Electroencephalography

Support for Electroencephalography (EEG) was developed as a BIDS Extension Proposal. Please see Citing BIDS on how to appropriately credit this extension when referring to it in the context of the academic literature.

Example datasets

Several example EEG datasets have been formatted using this specification and can be used for practical guidance when curating a new dataset.

EEG recording data

Template:

Legend:
  • For more information about filename elements (for example, entities, suffixes, extensions), follow the links embedded in the filename template.

  • Filename entities or directories between square brackets (for example, [_ses-<label>]) are OPTIONAL.

  • Some entities may only allow specific values, in which case those values are listed in <>, separated by |.

  • _<suffix> means that there are several (>6) valid suffixes for this filename pattern.

  • .<extension> means that there are several (>6) valid extensions for this file type.

  • [.gz] means that both the unzipped and gzipped versions of the extension are valid.

The EEG community uses a variety of formats for storing raw data, and there is no single standard that all researchers agree on. For BIDS, EEG data MUST be stored in one of the following formats:

Format Extension(s) Description
European data format .edf Each recording consists of a single .edf file. edf+ files are permitted. The capital .EDF extension MUST NOT be used.
BrainVision Core Data Format .vhdr, .vmrk, .eeg Each recording consists of a .vhdr, .vmrk, .eeg file triplet.
EEGLAB .set, .fdt The format used by the MATLAB toolbox EEGLAB. Each recording consists of a .set file with an OPTIONAL .fdt file.
Biosemi .bdf Each recording consists of a single .bdf file. bdf+ files are permitted. The capital .BDF extension MUST NOT be used.

It is RECOMMENDED to use the European data format, or the BrainVision data format. It is furthermore discouraged to use the other accepted formats over these RECOMMENDED formats, particularly because there are conversion scripts available in most commonly used programming languages to convert data into the RECOMMENDED formats.

We encourage users to provide additional metadata extracted from the manufacturer-specific data files in the sidecar JSON file.

Note the RecordingType, which depends on whether the data stream on disk is interrupted or not. Continuous data is by definition 1 segment without interruption. Epoched data consists of multiple segments that all have the same length (for example, corresponding to trials) and that have gaps in between. Discontinuous data consists of multiple segments of different length, for example due to a pause in the acquisition.

Note that for proper documentation of EEG recording metadata it is important to understand the difference between electrode and channel: An EEG electrode is attached to the skin, whereas a channel is the combination of the analog differential amplifier and analog-to-digital converter that result in a potential (voltage) difference that is stored in the EEG dataset. We employ the following short definitions:

  • Electrode = A single point of contact between the acquisition system and the recording site (for example, scalp, neural tissue, ...). Multiple electrodes can be organized as caps (for EEG), arrays, grids, leads, strips, probes, shafts, and so on.

  • Channel = A single analog-to-digital converter in the recording system that regularly samples the value of a transducer, which results in the signal being represented as a time series in the digitized data. This can be connected to two electrodes (to measure the potential difference between them), a magnetic field or magnetic gradient sensor, temperature sensor, accelerometer, and so on.

Although the reference and ground electrodes are often referred to as channels, they are in most common EEG systems not recorded by themselves. Therefore they are not represented as channels in the data. The type of referencing for all channels and optionally the location of the reference electrode and the location of the ground electrode MAY be specified.

Sidecar JSON (*_eeg.json)

For consistency between studies and institutions, we encourage users to extract the values of these fields from the actual raw data. Whenever possible, please avoid using ad hoc wording.

Those fields MUST be present:

Key name Requirement Level Data type Description
EEGReference REQUIRED string General description of the reference scheme used and (when applicable) of location of the reference electrode in the raw recordings (for example, "left mastoid", "Cz", "CMS"). If different channels have a different reference, this field should have a general description and the channel specific reference should be defined in the channels.tsv file.
SamplingFrequency REQUIRED number Sampling frequency (in Hz) of all the data in the recording, regardless of their type (for example, 2400). The sampling frequency of data channels that deviate from the main sampling frequency SHOULD be specified in the channels.tsv file.
PowerLineFrequency REQUIRED number or "n/a" Frequency (in Hz) of the power grid at the geographical location of the instrument (for example, 50 or 60).
SoftwareFilters REQUIRED object of objects or "n/a" Object of temporal software filters applied, or "n/a" if the data is not available. Each key-value pair in the JSON object is a name of the filter and an object in which its parameters are defined as key-value pairs (for example, {"Anti-aliasing filter": {"half-amplitude cutoff (Hz)": 500, "Roll-off": "6dB/Octave"}}).

Those fields SHOULD be present:

Key name Requirement Level Data type Description
CapManufacturer RECOMMENDED string Name of the cap manufacturer (for example, "EasyCap").
CapManufacturersModelName RECOMMENDED string Manufacturer's designation of the cap model (for example, "actiCAP 64 Ch Standard-2").
EEGChannelCount RECOMMENDED integer Number of EEG channels recorded simultaneously (for example, 21).

Must be a number greater than or equal to 0.
ECGChannelCount RECOMMENDED integer Number of ECG channels.

Must be a number greater than or equal to 0.
EMGChannelCount RECOMMENDED integer Number of EMG channels.

Must be a number greater than or equal to 0.
EOGChannelCount RECOMMENDED integer Number of EOG channels.

Must be a number greater than or equal to 0.
MISCChannelCount RECOMMENDED integer Number of miscellaneous channels not covered otherwise.

Must be a number greater than or equal to 0.
TriggerChannelCount RECOMMENDED integer Number of channels for digital (binary TTL) triggers or analog equivalents (TTL in volt). Corresponds to the TRIG channel type.

Must be a number greater than or equal to 0.
RecordingDuration RECOMMENDED number Length of the recording in seconds (for example, 3600).
RecordingType RECOMMENDED string Defines whether the recording is "continuous", "discontinuous", or "epoched", where "epoched" is limited to time windows about events of interest (for example, stimulus presentations or subject responses).

Must be one of: "continuous", "epoched", "discontinuous".
EpochLength RECOMMENDED number Duration of individual epochs in seconds (for example, 1) in case of epoched data. If recording was continuous or discontinuous, leave out the field.

Must be a number greater than or equal to 0.
EEGGround RECOMMENDED string Description of the location of the ground electrode (for example, "placed on right mastoid (M2)").
HeadCircumference RECOMMENDED number Circumference of the subject's head, expressed in cm (for example, 58).

Must be a number greater than 0.
EEGPlacementScheme RECOMMENDED string Placement scheme of EEG electrodes. Either the name of a standardized placement system (for example, "10-20") or a list of standardized electrode names (for example, ["Cz", "Pz"]).
HardwareFilters RECOMMENDED object of objects or "n/a" Object of temporal hardware filters applied, or "n/a" if the data is not available. Each key-value pair in the JSON object is a name of the filter and an object in which its parameters are defined as key-value pairs. For example, {"Highpass RC filter": {"Half amplitude cutoff (Hz)": 0.0159, "Roll-off": "6dB/Octave"}}.
SubjectArtefactDescription RECOMMENDED string Freeform description of the observed subject artifact and its possible cause (for example, "Vagus Nerve Stimulator", "non-removable implant"). If this field is set to "n/a", it will be interpreted as absence of major source of artifacts except cardiac and blinks.

These fields MAY be present:

Key name Requirement Level Data type Description
ElectricalStimulation OPTIONAL boolean Boolean field to specify if electrical stimulation was done during the recording (options are true or false). Parameters for event-like stimulation should be specified in the events.tsv file.

Must be one of: "true", "false".
ElectricalStimulationParameters OPTIONAL string Free form description of stimulation parameters, such as frequency or shape. Specific onsets can be specified in the events.tsv file. Specific shapes can be described here in freeform text.

Hardware information

Key name Requirement Level Data type Description
Manufacturer RECOMMENDED string Manufacturer of the equipment that produced the measurements.
ManufacturersModelName RECOMMENDED string Manufacturer's model name of the equipment that produced the measurements.
SoftwareVersions RECOMMENDED string Manufacturer's designation of software version of the equipment that produced the measurements.
DeviceSerialNumber RECOMMENDED string The serial number of the equipment that produced the measurements. A pseudonym can also be used to prevent the equipment from being identifiable, so long as each pseudonym is unique within the dataset.

Task information

Key name Requirement Level Data type Description
TaskName REQUIRED string Name of the task. No two tasks should have the same name. The task label included in the filename is derived from this "TaskName" field by removing all non-alphanumeric characters (that is, all except those matching [0-9a-zA-Z]). For example "TaskName" "faces n-back" or "head nodding" will correspond to task labels facesnback and headnodding, respectively. A RECOMMENDED convention is to name resting state task using labels beginning with rest.
TaskDescription RECOMMENDED string Longer description of the task.
Instructions RECOMMENDED string Text of the instructions given to subjects before the recording. This is especially important in context of resting state recordings and distinguishing between eyes open and eyes closed paradigms.
CogAtlasID RECOMMENDED string URI of the corresponding Cognitive Atlas Task term.
CogPOID RECOMMENDED string URI of the corresponding CogPO term.

Institution information

Key name Requirement Level Data type Description
InstitutionName RECOMMENDED string The name of the institution in charge of the equipment that produced the measurements.
InstitutionAddress RECOMMENDED string The address of the institution in charge of the equipment that produced the measurements.
InstitutionalDepartmentName RECOMMENDED string The department in the institution in charge of the equipment that produced the measurements.

Example *_eeg.json

{
  "TaskName":"Seeing stuff",
  "TaskDescription":"Subjects see various images for which phase, amplitude spectrum, and color vary continuously",
  "Instructions":"Your task is to detect images when they appear for the 2nd time, only then press the response button with your right/left hand (counterbalanced across subjects)",
  "InstitutionName":"The world best university, 10 Beachfront Avenue, Papeete",
  "SamplingFrequency":2400,
  "Manufacturer":"Brain Products",
  "ManufacturersModelName":"BrainAmp DC",
  "CapManufacturer":"EasyCap",
  "CapManufacturersModelName":"M1-ext",
  "EEGChannelCount":87,
  "EOGChannelCount":2,
  "ECGChannelCount":1,
  "EMGChannelCount":0,
  "MiscChannelCount":0,
  "TriggerChannelCount":1,
  "PowerLineFrequency":50,
  "EEGPlacementScheme":"10 percent system",
  "EEGReference":"single electrode placed on FCz",
  "EEGGround":"placed on AFz",
  "SoftwareFilters":{
    "Anti-aliasing filter":{
      "half-amplitude cutoff (Hz)": 500,
      "Roll-off": "6dB/Octave"
    }
  },
  "HardwareFilters":{
    "ADC's decimation filter (hardware bandwidth limit)":{
      "-3dB cutoff point (Hz)":480,
      "Filter order sinc response":5
    }
  },
  "RecordingDuration":600,
  "RecordingType":"continuous"
}

Note that the date and time information SHOULD be stored in the Study key file (scans.tsv). Date time information MUST be expressed as indicated in Units

Channels description (*_channels.tsv)

Template:

Legend:
  • For more information about filename elements (for example, entities, suffixes, extensions), follow the links embedded in the filename template.

  • Filename entities or directories between square brackets (for example, [_ses-<label>]) are OPTIONAL.

  • Some entities may only allow specific values, in which case those values are listed in <>, separated by |.

  • _<suffix> means that there are several (>6) valid suffixes for this filename pattern.

  • .<extension> means that there are several (>6) valid extensions for this file type.

  • [.gz] means that both the unzipped and gzipped versions of the extension are valid.

This file is RECOMMENDED as it provides easily searchable information across BIDS datasets. For example for general curation, response to queries, or for batch analysis. To avoid confusion, the channels SHOULD be listed in the order they appear in the EEG data file. Any number of additional columns MAY be added to provide additional information about the channels.

Note that electrode positions SHOULD NOT be added to this file, but to *_electrodes.tsv. Furthermore, the entries in *_electrodes.tsv and *_channels.tsv do not have to match exactly, as for example in the case of recording a single EOG channel from a bipolar referencing scheme of two electrodes, or a data channel originating from an auxiliary, non-electrode device. That is, in most cases *_electrodes.tsv will have more entries than *_channels.tsv. See the examples for *_channels.tsv below, and for *_electrodes.tsv in "Electrodes description".

The columns of the channels description table stored in *_channels.tsv are:

Column name Requirement Level Data type Description
name REQUIRED string Label of the channel.

Values in name MUST be unique.

This column must appear first in the file.
type REQUIRED string Type of channel; MUST use the channel types listed below. Note that the type MUST be in upper-case.

This column must appear second in the file.

For a list of valid values for this column, see the associated glossary entry.
units REQUIRED string Physical unit of the value represented in this channel, for example, V for Volt, or fT/cm for femto Tesla per centimeter (see Units).

This column must appear third in the file.
description OPTIONAL string Brief free-text description of the channel, or other information of interest.

This column may appear anywhere in the file.
sampling_frequency OPTIONAL number Sampling rate of the channel in Hz.

This column may appear anywhere in the file.
reference OPTIONAL string Name of the reference electrode(s). This column is not needed when it is common to all channels. In that case the reference electrode(s) can be specified in *_eeg.json as EEGReference).

This column may appear anywhere in the file.
low_cutoff OPTIONAL number or "n/a" Frequencies used for the high-pass filter applied to the channel in Hz. If no high-pass filter applied, use n/a.

This column may appear anywhere in the file.
high_cutoff OPTIONAL number or "n/a" Frequencies used for the low-pass filter applied to the channel in Hz. If no low-pass filter applied, use n/a. Note that hardware anti-aliasing in A/D conversion of all MEG/EEG electronics applies a low-pass filter; specify its frequency here if applicable.

This column may appear anywhere in the file.
notch OPTIONAL string Frequencies used for the notch filter applied to the channel, in Hz. If notch filters are applied at multiple frequencies, these frequencies MAY be specified as a list, for example, [60, 120, 180]. If no notch filter was applied, use n/a.

This column may appear anywhere in the file.
status OPTIONAL string Data quality observed on the channel. A channel is considered bad if its data quality is compromised by excessive noise. If quality is unknown, then a value of n/a may be used. Description of noise type SHOULD be provided in [status_description].

This column may appear anywhere in the file.

Must be one of: "good", "bad", "n/a".
status_description OPTIONAL string Freeform text description of noise or artifact affecting data quality on the channel. It is meant to explain why the channel was declared bad in the status column.

This column may appear anywhere in the file.
Additional Columns OPTIONAL n/a Additional columns are allowed if they are defined in the associated metadata file.

Restricted keyword list for field type in alphabetic order (shared with the MEG and iEEG modality; however, only the types that are common in EEG data are listed here). Note that upper-case is REQUIRED:

Keyword Description
AUDIO Audio signal
EEG Electroencephalogram channel
EOG Generic electrooculogram (eye), different from HEOG and VEOG
ECG Electrocardiogram (heart)
EMG Electromyogram (muscle)
EYEGAZE Eye tracker gaze
GSR Galvanic skin response
HEOG Horizontal EOG (eye)
MISC Miscellaneous
PPG Photoplethysmography
PUPIL Eye tracker pupil diameter
REF Reference channel
RESP Respiration
SYSCLOCK System time showing elapsed time since trial started
TEMP Temperature
TRIG Analog (TTL in Volt) or digital (binary TTL) trigger channel
VEOG Vertical EOG (eye)

Examples of free-form text for field description

  • n/a
  • stimulus
  • response
  • skin conductance
  • battery status

Example *_channels.tsv

See also the corresponding electrodes.tsv example.

name     type  units  description                     reference     status  status_description
VEOG     VEOG  uV     left eye                        VEOG-, VEOG+  good    n/a
FDI      EMG   uV     left first dorsal interosseous  FDI-, FDI+    good    n/a
Cz       EEG   uV     n/a                             REF           bad     high frequency noise
UADC001  MISC  n/a    envelope of audio signal        n/a           good    n/a

Electrodes description (*_electrodes.tsv)

Template:

Legend:
  • For more information about filename elements (for example, entities, suffixes, extensions), follow the links embedded in the filename template.

  • Filename entities or directories between square brackets (for example, [_ses-<label>]) are OPTIONAL.

  • Some entities may only allow specific values, in which case those values are listed in <>, separated by |.

  • _<suffix> means that there are several (>6) valid suffixes for this filename pattern.

  • .<extension> means that there are several (>6) valid extensions for this file type.

  • [.gz] means that both the unzipped and gzipped versions of the extension are valid.

File that gives the location of EEG electrodes. Note that coordinates are expected in cartesian coordinates according to the EEGCoordinateSystem and EEGCoordinateUnits fields in *_coordsystem.json. If an *_electrodes.tsv file is specified, a *_coordsystem.json file MUST be specified as well. The order of the required columns in the *_electrodes.tsv file MUST be as listed below.

Column name Requirement Level Data type Description
name REQUIRED string Name of the electrode contact point.

Values in name MUST be unique.

This column must appear first in the file.
x REQUIRED number Recorded position along the x-axis.

This column must appear second in the file.
y REQUIRED number Recorded position along the y-axis.

This column must appear third in the file.
z REQUIRED number or "n/a" Recorded position along the z-axis.

This column must appear fourth in the file.
type RECOMMENDED string Type of the electrode (for example, cup, ring, clip-on, wire, needle).

This column may appear anywhere in the file.
material RECOMMENDED string Material of the electrode (for example, Tin, Ag/AgCl, Gold).

This column may appear anywhere in the file.
impedance RECOMMENDED number Impedance of the electrode, units MUST be in kOhm.

This column may appear anywhere in the file.
Additional Columns OPTIONAL n/a Additional columns are allowed if they are defined in the associated metadata file.

*_electrodes.tsv files SHOULD NOT be duplicated for each data file, for example, during multiple runs of a task. The inheritance principle MUST be used to find the appropriate electrode positions for a given data file. If electrodes are repositioned, it is RECOMMENDED to use multiple sessions to indicate this.

Example *_electrodes.tsv

See also the corresponding electrodes.tsv example.

name   x        y          z        type     material
VEOG+  n/a      n/a      n/a      cup      Ag/AgCl
VEOG-  n/a      n/a      n/a      cup      Ag/AgCl
FDI+   n/a      n/a      n/a      cup      Ag/AgCl
FDI-   n/a      n/a      n/a      cup      Ag/AgCl
GND    -0.0707  0.0000   -0.0707  clip-on  Ag/AgCl
Cz     0.0000   0.0714   0.0699   cup      Ag/AgCl
REF    -0.0742  -0.0200  -0.0100  cup      Ag/AgCl

The acq-<label> entity can be used to indicate acquisition of the same data. For example, this could be the recording of electrode positions with a different electrode position recording device, or repeated digitization before and after the recording.

Coordinate System JSON (*_coordsystem.json)

Template:

Legend:
  • For more information about filename elements (for example, entities, suffixes, extensions), follow the links embedded in the filename template.

  • Filename entities or directories between square brackets (for example, [_ses-<label>]) are OPTIONAL.

  • Some entities may only allow specific values, in which case those values are listed in <>, separated by |.

  • _<suffix> means that there are several (>6) valid suffixes for this filename pattern.

  • .<extension> means that there are several (>6) valid extensions for this file type.

  • [.gz] means that both the unzipped and gzipped versions of the extension are valid.

A *_coordsystem.json file is used to specify the fiducials, the location of anatomical landmarks, and the coordinate system and units in which the position of electrodes and landmarks is expressed. The *_coordsystem.json is REQUIRED if the optional *_electrodes.tsv is specified. If a corresponding anatomical MRI is available, the locations of landmarks and fiducials according to that scan should also be stored in the *_T1w.json file which goes alongside the MRI data.

For disambiguation, we employ the following definitions for fiducials and anatomical landmarks respectively:

  • Fiducials are objects with a well defined location used to facilitate the localization of electrodes and co-registration with other geometric data such as the subject's own T1 weighted magnetic resonance head image, a T1 weighted template head image, or a spherical head model. Commonly used fiducials are vitamin-E pills, which show clearly in an MRI, or reflective spheres that are localized with an infrared optical tracking system.

  • Anatomical landmarks are locations on a research subject such as the nasion, which is the intersection of the frontal bone and two nasal bones of the human skull.

Fiducials are typically used in conjunction with anatomical landmarks. An example would be the placement of vitamin-E pills on top of anatomical landmarks, or the placement of LEDs on the nasion and preauricular points to triangulate the position of other LED-lit electrodes on a research subject's head.

General fields:

Key name Requirement Level Data type Description
IntendedFor OPTIONAL string or array The paths to files for which the associated file is intended to be used. Contains one or more BIDS URIs. Using forward-slash separated paths relative to the subject subdirectory is DEPRECATED. This identifies the MRI or CT scan associated with the electrodes, landmarks, and fiducials.

Fields relating to the EEG electrode positions:

Key name Requirement Level Data type Description
EEGCoordinateSystem REQUIRED string Defines the coordinate system for the EEG sensors.
See the Coordinate Systems Appendix for a list of restricted keywords for coordinate systems. If "Other", provide definition of the coordinate system in EEGCoordinateSystemDescription.

For a list of valid values for this field, see the associated glossary entry.
EEGCoordinateUnits REQUIRED string Units of the coordinates of EEGCoordinateSystem.

Must be one of: "m", "mm", "cm", "n/a".
EEGCoordinateSystemDescription RECOMMENDED, but REQUIRED if EEGCoordinateSystem is "Other" string Free-form text description of the coordinate system. May also include a link to a documentation page or paper describing the system in greater detail.

Fields relating to the position of fiducials measured during an EEG session/run:

Key name Requirement Level Data type Description
FiducialsDescription OPTIONAL string Free-form text description of how the fiducials such as vitamin-E capsules were placed relative to anatomical landmarks, and how the position of the fiducials were measured (for example, "both with Polhemus and with T1w MRI").
FiducialsCoordinates RECOMMENDED object of arrays Key-value pairs of the labels and 3-D digitized position of anatomical landmarks, interpreted following the "FiducialsCoordinateSystem" (for example, {"NAS": [12.7,21.3,13.9], "LPA": [5.2,11.3,9.6], "RPA": [20.2,11.3,9.1]}). Each array MUST contain three numeric values corresponding to x, y, and z axis of the coordinate system in that exact order.
FiducialsCoordinateSystem RECOMMENDED string Defines the coordinate system for the fiducials. Preferably the same as the "EEGCoordinateSystem". See the Coordinate Systems Appendix for a list of restricted keywords for coordinate systems. If "Other", provide definition of the coordinate system in "FiducialsCoordinateSystemDescription".

For a list of valid values for this field, see the associated glossary entry.
FiducialsCoordinateUnits RECOMMENDED string Units in which the coordinates that are listed in the field "FiducialsCoordinateSystem" are represented.

Must be one of: "m", "mm", "cm", "n/a".
FiducialsCoordinateSystemDescription RECOMMENDED, but REQUIRED if FiducialsCoordinateSystem is "Other" string Free-form text description of the coordinate system. May also include a link to a documentation page or paper describing the system in greater detail.

Fields relating to the position of anatomical landmark measured during an EEG session/run:

Key name Requirement Level Data type Description
AnatomicalLandmarkCoordinates RECOMMENDED object of arrays Key-value pairs of the labels and 3-D digitized locations of anatomical landmarks, interpreted following the "AnatomicalLandmarkCoordinateSystem" (for example, {"NAS": [12.7,21.3,13.9], "LPA": [5.2,11.3,9.6], "RPA": [20.2,11.3,9.1]}. Each array MUST contain three numeric values corresponding to x, y, and z axis of the coordinate system in that exact order.
AnatomicalLandmarkCoordinateSystem RECOMMENDED string Defines the coordinate system for the anatomical landmarks. See the Coordinate Systems Appendix for a list of restricted keywords for coordinate systems. If "Other", provide definition of the coordinate system in "AnatomicalLandmarkCoordinateSystemDescription". Preferably the same as the EEGCoordinateSystem.

For a list of valid values for this field, see the associated glossary entry.
AnatomicalLandmarkCoordinateUnits RECOMMENDED string Units of the coordinates of "AnatomicalLandmarkCoordinateSystem".

Must be one of: "m", "mm", "cm", "n/a".
AnatomicalLandmarkCoordinateSystemDescription RECOMMENDED, but REQUIRED if AnatomicalLandmarkCoordinateSystem is "Other" string Free-form text description of the coordinate system. May also include a link to a documentation page or paper describing the system in greater detail.

If the position of anatomical landmarks is measured using the same system or device used to measure electrode positions, and if thereby the anatomical landmarks are expressed in the same coordinates, the coordinates of the anatomical landmarks can be specified in electrodes.tsv. The same applies to the coordinates of the fiducials.

Anatomical landmarks or fiducials measured on an anatomical MRI that match the landmarks or fiducials during an EEG session/run, must be stored separately in the corresponding *_T1w.json or *_T2w.json file and should be expressed in voxels (starting from [0, 0, 0]).

Example *_coordsystem.json

{
  "IntendedFor":"bids::sub-01/ses-01/anat/sub-01_T1w.nii",
  "EEGCoordinateSystem":"Other",
  "EEGCoordinateUnits":"mm",
  "EEGCoordinateSystemDescription":"RAS orientation: Origin halfway between LPA and RPA, positive x-axis towards RPA, positive y-axis orthogonal to x-axis through Nasion,  z-axis orthogonal to xy-plane, pointing in superior direction.",
  "FiducialsDescription":"Electrodes and fiducials were digitized with Polhemus, fiducials were recorded as the center of vitamin E capsules sticked on the left/right pre-auricular and on the nasion, these are also visible on the T1w MRI"
}

Landmark photos (*_photo.<extension>)

Photos of the anatomical landmarks and/or fiducials.

Template:

Legend:
  • For more information about filename elements (for example, entities, suffixes, extensions), follow the links embedded in the filename template.

  • Filename entities or directories between square brackets (for example, [_ses-<label>]) are OPTIONAL.

  • Some entities may only allow specific values, in which case those values are listed in <>, separated by |.

  • _<suffix> means that there are several (>6) valid suffixes for this filename pattern.

  • .<extension> means that there are several (>6) valid extensions for this file type.

  • [.gz] means that both the unzipped and gzipped versions of the extension are valid.

Photos of the anatomical landmarks and/or fiducials are OPTIONAL. Please note that the photos may need to be cropped or blurred to conceal identifying features prior to sharing, depending on the terms of the consent given by the subject.

The acq-<label> entity can be used to indicate acquisition of different photos of the same face (or other body part in different angles to show, for example, the location of the nasion (NAS) as opposed to the right periauricular point (RPA).

Example *_photo.<extension>

Picture of a NAS fiducial placed between the eyebrows, rather than at the actual anatomical nasion: sub-0001_ses-001_acq-NAS_photo.jpg

placement of NAS fiducial